<会議発表論文>
Biomechanical Evaluation of Three Dimensionally Printed Passive Transfemoral Leg Socket Compliance with ISO 10328: A Comparison of Two Different Designs

作成者
本文言語
出版者
発行日
収録物名
開始ページ
終了ページ
会議情報
出版タイプ
アクセス権
Crossref DOI
権利関係
権利関係
概要 Malaysia’s need for 200,000 prosthetic legs demands efficient production processes and three-dimensional (3D) printing can potentially address this. Despite the advantages of pin-lock sockets in below...-knee prosthetics, research on their application in transfemoral amputations remains limited. This study designs and compares conventional and pin- lock prosthetic leg sockets for transfemoral amputees using 3D scanning, modeling, and printing. Conventional and pin- lock transfemoral leg sockets at 6 mm were designed using 3-Matic software. Their biomechanical properties were tested via ISO 10328 compression testing on the Instron 8874 machine. Results indicated that pin-lock sockets had higher average Young’s modulus (36.27%) and yield strength (43.53%) compared to conventional designs. However, ultimate compressive strength (UCS) showed no statistically significant difference between the designs (p > 0.05). Practical benefits like user stability and comfort might favor the pin-lock design. Future research should consider larger sample sizes and additional gait phases for comprehensive evaluation.続きを見る

本文ファイル

pdf 2024_p659 pdf なし 587 KB 729  

詳細

EISSN
レコードID
主題
助成情報
登録日 2024.11.12