<学術雑誌論文>
The Influence of Pore Size and Material Properties on Biomechanical Analysis of Parietal-Temporal Implant

作成者
本文言語
出版者
発行日
収録物名
開始ページ
終了ページ
出版タイプ
アクセス権
Crossref DOI
権利関係
概要 This study aimed to analyze the biomechanical behavior of the cranial implant by varying materials (titanium, hydroxyapatite, and zinc-hydroxyapatite) and pore size (600, 900, and 1200 μm) through a f...inite element method since there is a lack of study on this area. The cranial implant was reconstructed from the DICOM images, and the biomechanical behavior was evaluated in terms of Von Mises stress and deformation. The maximum Von Mises stress was primarily influenced by pore size in this study. Meanwhile, the material properties had the greatest influence on the implant’s deformation. Subsequently, the safety factor was considered, and titanium implant with a pore size of 900 μm showed excellent results.続きを見る

本文ファイル

pdf 750-758 pdf 1.56 MB 534  

詳細

PISSN
EISSN
レコードID
査読有無
主題
登録日 2022.01.05
更新日 2024.02.21

この資料を見た人はこんな資料も見ています