<departmental bulletin paper>
Characteristics of the Ne-based Gases as a Counter Filling Gas

Creator
Language
Publisher
Date
Source Title
Vol
Issue
First Page
Last Page
Publication Type
Access Rights
JaLC DOI
Related DOI
Related URI
Relation
Abstract Characteristics of gas multiplication and position sensing in Ne(neon)-based gases for radiation gas counters are investigated, and results are compared with those of Ar(argon)-based gases. The result...s show that with Ne-based gases is got about ten times larger gas gain compared with Ar-based gases at the same anode voltage, and comparable position resolution at about 200 volt lower high voltage. It is proposed that Ne-based gases are preferable comparing with Ar-based gases for proportiona1, and drift counter filling gases.show more