<departmental bulletin paper>
Theoretical Evaluation of the Detection Ratio of the Magnetization Due to the Geometrical Configurations on AC Loss Measurement with a Saddle-Shaped Pickup Coil

Creator
Language
Publisher
Date
Source Title
Vol
Issue
First Page
Last Page
Publication Type
Access Rights
JaLC DOI
Related DOI
Related URI
Relation
Abstract We theoretically investigated adequate geometrical configuration of a saddle-shaped pickup coil for accurate measurement of the AC losses in oxide superconducting taped wires with large anisotropy. De...fining the ratio of the apparent signal due to the magnetic moment in sample wires, which is detected by a saddle-shaped pickup coil, to the actual one as a detection ratio, we calculated the dependences of the detection ratio on the height, width and length of the straight and opening parts of a saddle-shaped pickup coil. We found out that it is possible to obtain the actual magnetization without compensation by supposing that the effective length of sample wires is equal to that of the straight part of a saddle-shaped pickup coil. Here we should consider that the signal detected by the opening part compensates for the shortage of the signal detected by straight part. Finally the recommendable dimensions of the saddle-shaped pickup coil were given for the measurement of the AC losses in currently developed oxide superconducting wires.show more

Hide fulltext details.

pdf p077 pdf 2.97 MB 357  

Details

PISSN
EISSN
NCID
Record ID
Peer-Reviewed
Subject Terms
Created Date 2015.06.03
Modified Date 2020.12.09

People who viewed this item also viewed