Carbon and Oxygen Stable Isotopic Measurements of Carbonate and Water Samples using Mass Spectrometer with Gas Bench

Creator
Language
Publisher
Date
Source Title
Vol
First Page
Last Page
Publication Type
Access Rights
JaLC DOI
Related DOI
Related URI
Relation
Abstract We verified the appropriate conditions and the precisions of stable isotopic analyses for carbonate and water samples using a mass spectrometer with Gas Bench. Standard deviation and reproducibility o...f the analyses are dependent on the amount of carbon dioxide obtained from samples, and are minimized when the amount of carbonate sample of 0.09-0.14 mg. Isotopic equilibrium is obtained in l hour at 50℃ after the reaction of carbonate and purified phosphoric acid. On these conditions, the reproducibility was±0.19‰ forδ^13C, and±0.22‰ for δ^18O(2SD). In the case of water samples, the appropriate sample amount was 0.75-1.0 mL for seawater, and 0.5 mL for calcareous freshwater that includes 3-4 mM of dissolved inorganic carbon. Reproducibility of ±0.04‰ (2SD) is gained for δ^13C after the reaction for 20 hours at 25℃. Oxygen isotope reaches the equilibrium in 24 hours after the reaction, which reveals the reproducibility of ±0.18‰(2SD). Standard deviation and reproducibility in any cases are within a required level for geoscientific researches on the marine and terrestrial specimens.show more