<学術雑誌論文>
Portable Neutron Detector Using Ce:LiCaAlF_6 Scintillator

作成者
本文言語
出版者
発行日
収録物名
開始ページ
終了ページ
出版タイプ
アクセス権
権利関係
関連DOI
関連URI
関連HDL
概要 We fabricated a prototype portable neutron detector using a Ce:LiCaAlF_6 (Ce:LiCAF) scintillator and evaluated its basic performance. The weight of the fabricated detector including the battery is 800... g, significantly less than that of the conventional portable neutron detector, which weighs more than 5 kg. The neutron sensitivity of the fabricated detector is evaluated to be 2.3 cps/(µSv/h), comparable to that of the conventional detector. The gamma-ray intrinsic detection efficiency is evaluated to be less than 10^<−6>. The gamma-ray sensitivity is sufficiently low under ordinary circumstances. The expected count rate of the error counting due to the pile-up effect of gamma-ray-induced events is confirmed to be less than the background count rate for an incident gamma-ray rate of less than 10^<4> gammas/s.続きを見る