<学術雑誌論文>
Magnetic Microbeads Trapping using Microfluidic and Permanent Magnet System

作成者
本文言語
出版者
発行日
収録物名
開始ページ
終了ページ
出版タイプ
アクセス権
Crossref DOI
権利関係
概要 In Lab-on-Chip (LoC) magnetic system, magnetic microbeads are used in selective binding and separation of targeted biological cells under influence of a magnetic field. Enhancing magnetic trapping eff...iciency is vital for further biological analyses. In this study, trapping efficiency of 4.5 μm diameter magnetic microbeads under permanent magnet configurations and varying microfluidic flow rates has been investigated. In the experimental studies conducted, neodymium iron boron (NdFeB) permanents magnets and polydimethylsiloxane microfluidics with trapping chamber design were used as a LoC magnetic system. At 5 μl/min microfluidics flow, trapping efficiency of single magnet configuration was 92.06 %. Improved trapping efficiency of 95.24 % was observed with double magnet configuration. On the other hand, the trapping efficiency at greater microfluidics flow rate of 20 μl/min were only 20.63 % and 15.87 % for single and double magnet configuration respectively. Trapping efficiency is greater for double permanent magnets configuration and at the slowest microfluidics flow rate.続きを見る